Quality Control Calibrator Dose II




CRC-35R "Radioisotopes dose calibrator Owner's manual" - Capintec Inc.




Press TEST to enter the menu and choose the procedure DAILY to begin the test day. This procedure consists of the following controls:

    Size
  1. Fund - Background
  2. Test System - System Test
  3. Test Accuracy - Accuracy test
  4. Test Constance - Constancy Test

1. Size of the Fund - Background

  • This procedure determines the value of the environmental fund that is automatically subtracted for each measurement. It 'so important during the implementation of the measure there are no radioactive corgenti in the cockpit or in the vicinity.
  • Press BEGIN and await the outcome of the measure Please Wait
  • if at the end of the measure shows:
  • OK : take notadel value of the fund and prepare the F3 key
  • HIGH : nnotadel take the fund's value and repeat the measurement by pressing a button
  • TOO HIGH: press F3
  • Pay attention to the presence of radioactive sources in the vicinity and repeat the measurement. Where the value of the fund does not fall within normal range, notify the Health Physics


2. TEST SYSTEM - System test

  • E 'important that during the implementation of the measure there are no radioactive sources in or near the cockpit.
  • Press BEGIN and await the outcome of the measure PLEASE WAIT
  • Record the voltage measured: VOLTS: V
  • of its deviation from the nominal value: DEVIATION:%
  • and the outcome of the test: OK - FAIL
  • Prepare the ENTER key



3. TEST ACCURACY - Accuracy test

    Lower in the cockpit
  • the low-activity source calibration (133Ba content in shielding BLACK and serial number: s8107015 or other) and press F4
  • Record the value of assets measured ACCURACY mCi
  • its rated deviation: DEVIATION%
  • press the button to save the current data F3
If the message appeared ERROR write down the message and notify the Health Physics.



4. TEST CONSISTENCY - Constancy Test

  • Lower in the cockpit of the calibration source 133Ba and press F4
  • Record the value of assets measured CONSTANCY: mCi
  • Of his departure from value DEVIATION:%
  • Press SAVE or F3 to save the current data
  • Press DONE or F1 to exit test
  • Press F1 key to end the controls
  • Select Radionuclide measurement used in the ordinary course